Overviews
ELICO’s SL- 194, AAS Spectrophotometer is fully automated and delivers a new level of flexibility and simplicity. It can work as standalone mode for routine analysis and can also be fully PC Controlled. The selected element lamp is automatically aligned in the built-in 6/8 lamp turret which automatically recognizes coded HCL lamps and loads their operating parameters. The instrument provides many features in terms of performance and saftey never before available on low-cost AAS
Features
SALIENT FEATURES
- Built in 4 Line LCD Graphical Display
- Automated (6/8) lamp turret compatible to coded & non-coded HCL Lamps with independent Power Supply for warm-up
- High Energy True Double Beam Optics with Narrow Beam, fully sealed and Quartz over coated mirrors
- Background correction with D2 lamp source
- Integral Auto gas control system with LCD display
- Automatic Ignition and Flame Control with Full Saftey Interlocks
- Automatic Lamp Alignment
- Burner Movement in all three directions – height, rotation and lateral.
- Modes of operation include Absorbance, Emission, Concentration and Standard Addition
- Optional : 21 CFR Part 11 PC Software, Auto Sampler, Hydride and Cold Vapour Generator, Graphite Furnance.
Specification
TECHNICAL SPECIFICATIONS
SPECTRAL | |
Range | 185 to 900 nm |
Bandwidth | Automatic variable slit from 0 to 2.5 nm continuously variable with increment of 0.1 nm |
Readability | 0.1 nm |
Accuracy | ± 0.5 nm |
Repeatability | ± 0.2 nm |
PHOTOMETRIC | |
Range | 0 to 2 Abs |
Accuracy | ± 0.010 Abs at 1.0 Abs |
Integration Time | 0.04 to 99 sec |
LIGHT SOURCE |
6/8 Lamps Automatic Turret with Independent Power Supply compatible to coded and non coded HC Lamps. |
MONOCHROMATOR |
Czerney Turner 1/3 m with 1800 lines/mm Holographic Grating and Hard Coded Optics. |
BACK GROUND CORRECTION |
Deuterium Lamp (D2) |
BURNER |
Titanium construction, Premix Laminar Flow100mm burner for Air – Acetylene (C2H2) |
Optional – 50 mm burner for Nitrous Oxide (N O) – 2 Acetylene (C2H2) |
The Spray Chamber is made of Inert Material .i.e., PTFE (Teflon) fitted with adjustable impact bead Aerosol |
NEUBULIZER | |
Made of fully Inert Material with 90% Platinum and 10% Iridium Alloy Capillary | |
SAMPLE FEED RATE | 3 to 8 ml/min (Adjustable) |
DETECTOR |
Wide Range Photo Multiplier Tube (PMT) |
FLOW CONTROL |
Integral Auto Gas Control System with LCD Display, Keyboard and all Fault Indicators with all safety features |
Automatic Flame Changeover, Auto Ignition and Shut Off |
Optimization of Flame Conditions for each Element with programmable Flame Control |
SAFETY FEATURES |
Full Safety Interlocks including Pressure Sensors on both lines, Flame failure Detection, Burner Presence and Recognition, Nebulizer Presence and Drain Sensor, Power Failure Protection and Emergency Flame Shut Off Switch and Flash Back Arrester on the Cylinder |
SYSTEM UPGRADATION |
Instrument can be upgraded to Auto Sampler, Hydride and Cold vapour Generator |
ACCESSORIES SUPPLIED | |
With Instrument | 100 mm burner for Air – C2H2 , Nebulizer with SS Capillary, Back Ground Correction with D2 Lamp, Oil free Air Compressor with 100% Dustfree Air Filter, Acetylene (C2H2) regulator with Flashback Arrester. |
At extra Cost | 50 mm burner for N2O2 – C2H2 , Nebulizer with Pt – Ir Capillary, N2O Regulator, Nitrogen Regulator, HC Lamps for desired Elements coded or non coded, Hydride and Cold vapour Generator, Printer, Computer, PC Software, Exhaust Fumehood Chamber, Servo Controlled Voltage Stabilizer of suitable rating, Acetylene Gas Cylinder, Nitrous Oxide Gas Cylinder and Nitrogen Gas Cylinder. |
Brochure