The U6000AT+ adds a membrane desolvation module to the standard U5000AT+ Ultrasonic Nebulizer for improved reduction of solvent-based interferences, particularly for volatile organic solvents (both polar and non-polar types).
Use with ICP-AES: For ICP-AES the U6000AT+ is used primarily for the introduction of volatile organic solvents. These solvents can cause a number of serious problems: loss of plasma, plasma instability, high carbon emission background, and carbon buildup on the ICP torch.
Use with ICP-MS: For ICP-MS the U6000AT+ can be used for both volatile organics and aqueous samples
Higher Nebulization Efficiency: The piezoelectric transducer of the U6000AT+ converts more of the liquid sample into a usable aerosol, with an efficiency of 10-15%. With the membrane module connected, the result is a 2x to 5x enhancement of analyte signal vs. a standard pneumatic nebulizer.
Reduced Solvent-Based Interferences: The inert, fluoropolymer membrane desolvator reduces solvent based interferences from water and organic vapor. For example, the % CeO/Ce is typically <0.05% for water-based samples with the U6000AT+. Carbon-based inerferences such as 12C2+ are also greatly reduced, allowing detection of Mg at m/z = 24.
Improved Detection Limits: the combination of signal enhancement and reduced interferences enables lower instrument detection limits for both ICP-AES and ICP-MS.
Easy Setup: All U6000AT+ units are provided with an interface kit for quick and easy connection to the host ICP-AES or ICP-MS. The kit includes a nebulizer gas line and a sample out line with the appropriate ICP torch adapter. In addition, a rinse kit is included for cleaning the membrane desolvator.
Compact Design: The U6000AT+ consists of the standard U5000AT+ Ultrasonic Nebulizer and a modular membrane desolvator box. The U5000AT+ is simply placed on top of the desolvator box, conserving valuable benchtop space; the footprint of the U6000AT+ is 35.6 cm W x 34.9 cm D.
Flexibility: The membrane desolvator module can be easily disconnected from the ultrasonic nebulizer if membrane desolvation is not necessary for a particular application. This feature allows stand-alone operation of the ultrasonic nebulizer.
Optional Blend Gas Accessory: An optional blend gas accessory, the CETAC BGX-100, is available to help prevent carbon build-up on the ICP torch and ICP-MS sampling cones during extended introduction of volatile organic solvents.
Specification Title | Specification Details |
---|---|
Sample Uptake Rate | 0.5 – 2.5 ml/min |
Nebulizer Gas Flow | 0.5 – 1.5 L/min |
Heater Temperature | 120ºC to 160ºC |
Cooler Temperature | -20ºC to +10ºC |
Membrane Oven Temperature | 80ºC to 160ºC |
Voltage Ultrasonic Nebulizer | 100-120 VAC, 50/60Hz, 4.5A 220-240 VAC, 50/60Hz, 2.5A |
Voltage Membrane Desolvator | 100-240 VAC, 50/60Hz, 4A/2/A |
Height | 38.7 cm (15.25 in.) |
Width | 35.6 cm (14 in.) |
Depth | 34.9 cm (13.75 in.) |
Weight | 13.6 kg) (29.9 lbs) |
Warranty | 12 month limited |
- Polar organic solvents (ex. isopropyl alcohol)
- Non-polar organic solvents (ex. naphtha, toluene)
- Water-based samples for reduction of oxide and hydride interferences